The PET-CT Phantom™ includes internal structures (three rods and six spheres) which, when imaged with both modalities, can demonstrate how accurately the two image sets are aligned. In addition, a single sample of radioactive water is attenuated by water, bone and CT contrast material (as well as air only) to determine how accurately the CT-based PET attenuation correction works.
The Phantom is used for:
- Acceptance testing of PET/CT and SPECT/CT systems
- Routine quality evaluation of PET/CT and SPECT/CT systems
- Evaluation of new image fusion software
- Evaluation of new attenuation correction algorithms
- Aluminium tubes are for registration
- The outer 2" OD micro cylinder is for comparing attenuation region to non attenuation region
- The 6" ring is for contrast solution
- Research
| Interior length of phantom | 180 mm |
| Fillable spheres (5) inner diameter | 10, 13, 17, 22 and 28 mm |
| Distance from sphere plane to inside wall | 70 mm |
| Volume of empty cylinder | 9.7 litres |
| Main Cylindrical insert dimension | Outside diameter: 51 mm, Length: 180 mm |
| Top Cylinder | Cylinder outside diameter: ~5.1 cm, Cylinder inside diameter: ~4.cm, Cylinder inside height: ~8.2 cm, Cylinder outside height: ~12.0 cm, Volume of empty cylinder: 408 cm2 |
| Three Aluminium Tubes | One 5 inch long: ~1.7 cc, Two 7 inch (ea): ~ 2.5 cc |
| Stepped Bone Ring | Pre-filled with liquid bone composition, not to be opened. |
| Bone Ring Volumes | Outer volume: 15.6 cu. inch: ~256 cc, Inner volume: 6.7 cu. inch: ~110 cc |